Overview

Philips Azurion System

Health Canada – RA 73421

Issue:  Philips Azurion System may unexpectedly lose X-ray functionality. Due to a software issue, a mechanism that is present in the system to manage the number and size of log trace files does not function properly. Without this mechanism, the log trace files created by the system (e.g., at start, during use) may occupy the full disk capacity of the Philips Azurion system. When the full disk capacity is reached, X-ray functionality will cease to be available without an advance warning to the user.

Please visit for more information:  https://recalls-rappels.canada.ca/en/alert-recall/philips-azurion-system?utm_source=gc-notify&utm_medium=email&utm_content=en&utm_campaign=hc-sc-rsa-22-23