IEC 61000-4-21:2011
Electromagnetic compatibility (EMC) - Part 4-21: Testing and measurement techniques - Reverberation chamber test methods
Product Details
- In Clause 8, the use and specifications of E-field probes for application to reverberation chambers has been added.
- In Annex A, additional guidance and clarifications on the use of reverberation chambers at relatively low frequencies of operation (i.e., close to the lowest usable frequency of a given chamber) are given, and its implications on the estimation of field uncertainty are outlined.
- In Annex B, symmetric location of the field probes when the chamber exhibits cylindrical symmetry has been disallowed, as such placement could otherwise yield a false indication of field uniformity and chamber performance at different locations.
- Annex C now contains more quantitative guidance on the setting of the maximum permissible stirring speeds that warrant quasi-static conditions of operation for chamber validation and testing.
- In Annex D, a requirement for the EUT and equipment not to occupy more than 8 % of the total chamber volume in immunity testing has been added.
- Annex E has been extended with further guidance on the value of EUT directivity to be used in the estimation of radiated power and field.
- In Annex I, some clarifications on antenna efficiency measurements have been added.
- A new Annex K has been added that covers measurement uncertainty in reverberation chambers.