IEC 61010-2-033:2023
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-033: Particular requirements for hand-held multimeters and other meters for domestic and professional use, capable of measuring mains voltage
Product Details
a) Clause 2, all normative references have been dated and new normative references have been added;
b) 4.4.2.101 is a new subclause about surge protective devices;
c) Subclause 6.6.101 modifies 6.6.101 and 6.6.102 of previous edition:
1) in 6.6.101.1, insulating material of group I may be allowed for determination of creepage distances of measuring circuit terminals;
2) in 6.6.101.2, Clearances and creepage distances up to 3 000 V for measuring circuit terminals in unmated position have been defined;
3) in 6.6.101.3, requirements for measuring circuit terminals in partially mated position have been specified;
4) in 6.6.101.4, requirements for measuring circuit terminals in mated position have been specified;
5) Subclause 6.6.101.5 replaces 6.6.102;
d) Subclause 6.101 replaces 6.9.101 of the previous edition with modifications;
e) 9.101 is a new subclause to consider the protection of measuring circuits against the spread of fire and arc flash;
f) in 9.101.2, relocation of 101.3 of previous edition;
g) in 9.101.3, relocation of 101.4 of previous edition;
h) in 101.3, relocation of Clause 102 of previous edition;
i) in K.2.1, another method for determination of clearances of secondary circuits is proposed;
j) in K.3.2, new Table K.15 and Table K.16 for clearance calculation;
k) Clause K.4 of the previous edition has been deleted;
l) Subclause K.101.4 has been reviewed;
m) Table K.104 of the previous edition has been deleted;
n) Annex AA: Figure AA.1 has been redesigned;
o) Annex EE: addition of a new informative annex for determination of clearances for the purposes of Table 101.