IEC 61010-2-030:2023
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-030: Particular requirements for equipment having testing or measuring circuits
Product Details
a) in 1.2.1, requirements for protection against hazards which could occur from reading a voltage have been added to the scope;
b) Clause 2, all normative references have been dated and new normative references have been added;
c) in 4.3.2.5, requirements for power supply have been modified;
d) in 4.3.2.6, requirements for inputs/outputs have been modified;
e) in 4.4.2.101, a new subclause about surge protective devices has been added;
f) in 5.1.5.101.2, minimum ratings for voltage of measuring terminals are required;
g) Subclause 6.6.101 modifies 6.6.101 and 6.6.102 of previous edition:
1) in 6.6.101.1, insulating material of group I may be allowed for determination of creepage distances of measuring circuit terminals;
2) In 6.6.101.2, clearances and creepage distances up to 3 000 V for measuring circuit terminals in unmated position have been defined;
3) in 6.6.101.3, requirements for measuring circuit terminals in partially mated position have been specified;
4) in 6.6.101.4, requirements for measuring circuit terminals in mated position have been specified;
5) Subclause 6.6.101.5 replaces 6.6.102;
h) Subclause 9.101 to consider the protection of measuring circuits against the spread of fire and arc flash has been added and Table 102 has been replaced by Table K.101;
i) in 9.101.2, relocation of 101.3 of previous edition;
j) in 9.101.3, relocation of 101.4 of previous edition, extension to measurement category II and reference to IEC 61000-4-5 for tests;
k) in 14.101, relocation of 14.102 and 14.101 of previous edition has been removed;
l) in 101.3, relocation of 101.5 of previous edition, and more requirements added against hazard occurring from reading a voltage value;
m) in K.2.1, another method for determination of clearances of secondary circuits is proposed;
n) in K.3.2, new Table K.15 and Table K.16 for clearance calculation;
o) in K.101.4.1, new Table K.103 and Table K.104 replace Table K.102, Table K.103 and Table K.104;
p) in K.101.4, the subclause has been reviewed; Tables and tests for solid insulation have been modified; Table K.105 replaces Table K.9;
q) Table K.101, replacement of Table K.106;
r) Clause K.4, redraft of the clause to propose a method for determination of Ut for circuits which reduce transient overvoltages;
s) Annex AA: Figure AA.1 has been redesigned;
t) Annex EE: addition of a new informative annex for determination of clearances for Table 101.
It has the status of a group safety publication in accordance with IEC Guide 104.