Codes & Standards - Purchase
IEC PAS 63023:2016
Medical electrical system - Input interface for haemodialysis equipment for use of external alarming device
SKU: iec_024041_052368
Published by IEC
Publication Year 2016
1.0 Edition
12 pages
Product Details
IEC PAS 63023:2016 establishes a unique input interface for connection of an external alarming device to haemodialysis equipment. The input interface of the haemodialysis equipment is designed as a simple solution, which takes a single fault condition of the input interface into account, to stop the extracorporeal blood flow in case of needles slipping out from the fistula or graft detected by the external alarming device.